BioLogics, Inc.
  OMNICON ZONE READER
 


OMNICON Zone Reader
Zone Reader Pharmacopoeia Methods
USP Method
EP Method
SM Method
Zone Reader Peni Cylinders & Dispenser

 
OMNICON Zone Reader
 

European Pharmacopoeia (EP) and British Pharmacopoeia (BP) Methods
Performs all calculations in accordance to the procedures defined in the EP / BP Methods, sometime referred to as the 3+3 Method.

Plate Reading
The OMNICON Zones Reader reads each 100 mm petri dish in less than 3/10th of a second, measuring the diameter and shape of the zone. The shape value is used to exclude zones which are oval due to spills or irregular growth.

21 CFR, Part 11, Electronic Records and Signatures
The OMNICON Zones Reader is compliant, providing three levels of security, audit trails and data encryption.

Assay Layout

BP EP Antibiotic Potency Testing, OMNICON Zone Reader

 

6 zones per plate
Standard Concentrations - S1, S2 and S3
Preparations - P1, P2 and P3

The software allows the Analyst to specify a minimum of 4 plates and maximum of 10 plates per assay.

The Analyst defines the concentration of the standards and preparations, and the assumed potency, the software then measures the zone diameters and performs all the calculations and presents the data on the "Independent Assay" report.

The software also provides the Analyst with the ability to generate a Combination Analysis" report by selecting Independent assay previously analyzed.


 

References
United States Pharmacopoeia (USP)
US-FDA Code of Federal Regulations (CFR)
British Pharmacopoeia (BP)
European Pharmacopoeia (EP)

OMNICON is a registered trademark of BioLogics, Inc.

 
 